Description
1. Semiconductor Device Analyzer Mainframe (B1500A ):
– Ground unit (GNDU):
+ Output voltage: 0 V ±100 µV
+ Maximum sink current: ±4.2 A
– Data storage: SSD, DVD-R drive
– Interfaces: GPIB, interlock, USB (USB 2.0, front 2, rear 2), LAN (1000BASE-T/100BASE-TX/10BASE-T), trigger in/out, digital I/O, VGA video output
2. MFCMU (multi frequency capacitance measurement unit) Module
– Measurement parameters: Cp-G, Cp-D, Cp-Q, Cp-Rp, Cs-Rs, Cs-D, Cs-Q, Lp-G, Lp-D, Lp-Q, Lp-Rp, Ls-Rs, Ls-D, Ls-Q, R-X, G-B, Z-q, Y-q
– Frequency:
+ Range: 1 kHz to 5 MHz
+ Resolution: 1 mHz
+ Accuracy: ±0.008%
– Output signal level:
+ Range: 10 mVrms to 250 mVrms
+ Resolution: 1 mVrms
+ Accuracy: ±(10.0% + 1 mVrms) at the measurement port of the MFCMU
3. High power source/monitor unit (HPSMU):
– Range up to 200 V/1 A with 4-quadrant operation
– Minimum measurement resolution 10 fA/2 µV
4. High resolution source/monitor unit (HRSMU)
– Range up to 100 V/0.1 A with 4-quadrant operation
– Minimum measurement resolution 1 fA/0.5 μV
If you have extra question or have some projects having this product, please feel free to reach out to us!